Rui Zhao wrote:
> I wonder if somebody can give me some
>
> suggestions/comments that if the poor diffraction can be caused by the
> high
> solvent content or the intrinsic flexbility is likely to be the real
> cause? If the high
> solvent content does contribute to the poor diffraction, did anybody
> have
> experience/tricks to overcome this problem?
> Thank you very much for your help. Rui.
Well, maybe this help you:
1 - try to collect data from very fresh crystals (growing time sometimes
help us to distinguish between slightly different molecule conformations,
leading to better diffracting crystals);
2 - maybe you have problems while flash-cooling your crystals (remember,
not only ice is an important aspect in this issue...). Be sure that your
protocol is a good one (see details in Garman, E. F. & Schneider, T. R.
(1997), J Appl Cryst, 30: 211-237). You can check this doing a room
temperature data collection. As another approach, you can try crystal
annealing (Harp, J. M., et al. (1998), Acta Cryst. D54: 622-628) .
Good luck,
Ricardo
--
===========================================================